Ürünler

Emisyon Tomografisi Performans Fantomu

Phantom for performance evaluation of small animal systems or ultra-high resolution ECT systems.

It can be used to evaluate: center-of-rotation error, non-uniformity artifacts and spatial resolution. Available in two sizes for FOV at least 45mm or 75mm.

Technical data (can be modified to customer specifications):

  • at least 75mm FOV version:
    • main cylinder:
      • outside cylinder diameter: 80 mm
      • inside cylinder diameter: 74 mm
      • inside cylinder height: 65 mm
      • cold rods insert:
      • rod diameters: 1.2, 1.6, 2.4, 3.2, 4.0 and 4.8 mm
      • height of rods: 34 mm
  • at least 45mm FOV version:
    • main cylinder:
      • outside cylinder diameter: 50 mm
      • inside cylinder diameter: 45 mm
      • inside cylinder height: 65 mm
      • cold rods insert:
      • rod diameters: 1.2, 1.6, 2.4, 3.2, 4.0 and 4.8 mm
      • height of rods: 34 mm
  • optional carrying case

Product features:

  • Complies with:
    • NEMA Standards Publication (NU 1-2001) Performance Measurements of Scintillation Cameras
    • AAPM Report No. 9 - Computer Aided Scintillation Camera Acceptance Testing
    • AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
    • ACR–SNM (Res. 5 – 2011) technical standard for diagnostic procedures using radiopharmaceuticals
  • CE certified
  • the Manual provides detailed guidelines for carrying out each test, results assessment and registratio