Phantom for NM and PET systems performance evaluation (collimator, artifacts, calibration, reconstruction parameters). It can be used to evaluate, for example: center-of-rotation error, non-uniformity artifacts, changes of radius-of-rotation on spatial resolution, reconstruction filters on spatial resolution, attenuation and scatter compensation.
Technical data (can be modified to customer specifications):
- main cylinder:
– inside cylinder diameter: 206mm
– inside cylinder height: 186mm
– cylinder wall thickness: 7mm
- cold rods insert:
– rod diameters: 4.8, 6.4, 7.9, 9.5, 11.1 and 12.7 mm
– height of rods: 88mm
- cold spheres:
– solid sphere diameters: 9.5, 12.7, 15.9, 19.1, 25.4 and 31.8mm
– height of the center of the spheres from the base plate: 127mm
- optional PET Lid with cylindrical samples:
– refillable thin-walled cylinders, diameters: 8, 12, 16 and 25mm
– water filled cylinder diameter: 25mm
– air filled cylinder diameter: 25mm
– PTFE solid cylinder diameter: 25mm
– cylinder height: 38mm
- heavy duty carrying case
Product features:
- Complies with:
– NEMA Standards Publication (NU 1-2001) Performance Measurements of Scintillation Cameras
– AAPM Report No. 9 - Computer Aided Scintillation Camera Acceptance Testing
– AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
– ACR–SNM (Res. 5 – 2011) technical standard for diagnostic procedures using radiopharmaceuticals
- CE certified
- the Manual provides detailed guidelines for carrying out each test, results assessment and registration